Twin anode X-ray source
Page 2 of 6
Start-Up Procedure
-
Verify the main chamber base pressure is below 5x10-9 mbar.
-
Turn on the X-ray source power supply and wait at least 10 minutes for electronics thermal stabilization.
-
Open the water cooling circuit and set the water pressure to aprox. 4.8 bars.
-
Open drain valve first.
-
Slowly open the source valve until the pressure reaches approximately 4.8 bars.
Water cooling |
 |
-
Verify all interlock LEDs are green.
Power supply, Interlocks |
 |
-
Start Avantage software. Select OK when a licence window appears.
-
Navigate to the "Instrument" tab (A). The Xray Gun Control panel will appear on the top.
-
If the panel displays "Device OK" (B), software control is ready.
-
Expand the control panel by clicking the arrow on its right edge.
X ray Gun Control Panel |
 |
-
Select 15W from the Spot/Power list in the X-Ray Gun Control panel (C).
-
Monitor the main chamber pressure.
-
Press the Apply button (D). The gun ramping up procedure will start. Wait for the "Device OK" (B) message and monitor the pressure, which should remain below 5x10-9 mbar. (Initial pressure spikes up to 1x10-8 mbar mbar are normal.)
-
Once the pressure stabilizes (decreases), gradually increase the power to 100W, 200W, and finally 300W.
-
Wait at least 10 minutes for X-ray intensity to stabilize.