Twin anode X-ray source
Page 3 of 6
Shutdown Procedure
-
In the X-Ray Gun Control panel, press the Shutdown button and wait for the "Device OK" message to reappear.
-
Verify that the red "HV ON" indicator on the power supply is off.
-
Maintain water cooling for at least 10 minutes.
-
Close the water cooling circuit.
-
Slowly close the source valve first..
-
Then, close the drain valve.
-
Pull out the X-ray gun. Raise the linear transfer until the top red line mark is visible.
-
Exit Avantage software.