Twin anode X-ray source

Binding energy scale calibration


Pass energy 100 eV, A21 mode:
Convert kinetic energies to binding energies by subtracting the kinetic energy from the Fermi edge positions listed in the following table.


Anode Photon energy (eV) Fermi edge KE (eV) Resolution (eV)
Mg Kα 1253.6 1248.67 0.85
Al Kα 1486.6 1481.72 1.0